PrinCE with LED and Laser-Induced Fluorescence detectors

zetalif-440x280.png

LASER Induced Fluorescence detectors are extremely powerful, typically capable of emitting light of a very well defined wavelength with an intensity of 1-40 mW. Thus the maximum possible number of molecules can be excited, resulting in the best possible limits of detection.

For fluorescent compounds, concentration limits of detection down to 10-12 mol/l have been reported. Bearing in mind that typically about 10 nl of sample is injected, this equals about 10-20 mol! LIF detection can be more sensitive than any other commercially available detector, even if the compound is not fluorescent itself. Of course a suitable label should be available.

Comparable with the LIF, LED is from 10–100 times greater than conventional fluorescence and up to 100,000 times more sensitive than UV detection. Low noise and interference molecules will be detected. Only labeled or native fluorescence can be measured.


The CE-cell can be placed inside the temperature controlled capillary compartment. The detector can be used with any PrinCE and PrinCE Next system.

Specifications — LIF detector

Specifications

LIF Details

Available Laser Wavelengths

266, 325, 355, 410, 442, 473, 488, 514, 532, 594, 633, 638, 642 & 780 nm.

Emission Filter:

The emission filters depend on the wavelength and generally contain highpass filters

LIF Capillaries

OD 365 μm; ID from 20 to 320 μm

Cell Detection Volume

The cell is a small part of the capillary and depends on the capillary ID, for example 0.4 nL with a 50 μm ID.

Signal Outputs

- External Ethernet for Network communications.

- External Event (detection of a relay state) function. Start/stop command port for an external event command (relay opened or closed).

- Analog output: Processed: 0-1V(DC) for range 0-50 RFU. 0-100mV(DC) for range 0-5 RFU.

Power Requirements

100-240 VAC, 47/63 Hz, 1.5 A

Dimensions and Weight

43.0 cm / 16.9”(H) x 23.0 cm / 9.1”(W) x 34.0 cm / 13.33” (D). 12 kg / 26.4 lbs

 

Specifications — LED detector

Specifications

LED Details

Available LED Wavelengths

450, 480, 530 nm

Emission Filter:

The emission filters depend on the wavelength and generally contain highpass filters

LIF Capillaries

OD 365 μm; ID from 20 to 320 μm

Cell Ddetection Volume

The cell is a small part of the capillary and depends on the capillary ID, for example 0.4 nL with a 50 μm ID.

Signal Outputs

- External Ethernet for Network communications.

- External Event (detection of a relay state) function. Start/stop command port for an external event command (relay opened or closed).

- Analog output: Processed: 0-1V(DC) for range 0-50 RFU. 0-100mV(DC) for range 0-5 RFU.

Power Requirements

100-240 VAC, 47/63 Hz, 1.5 A

Dimensions and Weight

43.0 cm / 16.9”(H) x 23.0 cm / 9.1”(W) x 34.0 cm / 13.33” (D). 12 kg / 26.4 lbs

 